Vertical autoclaves Classic-Line: VQ-series

Best value for money with a vacuum function

Home Autoclaves Vertical autoclaves: Classic-Line Vertical autoclaves Classic-Line: VQ-series

Features at first glance:

  • Certified according to the Pressure Equipment Directive 2014/68/EU
  • Available with 71 liters of effective usable volume (chamber volume 82 litres)
  • Can also be fitted with exhaust air filtration on request for laboratories with biosafety level 2 or higher
  • Sterilisation temperatures up to 135°C
  • Packed with features as standard – comprehensive equipment for your daily laboratory work, such as:
    • Media temperature sensor for sterilising liquids
    • Vacuum function for sterilising porous materials, hollow items, textiles and more
    • Automatic water supply – can be switched on and off as required
    • Various preset and two individually programmable autoclaving programmes
    • Printer and SD card slot for documenting your sterilisation cycles
    • Castors for increased mobility of the unit
  • Electrical, mechanical and constructional factors guarantee the highest level of user protection
  • Techn. data VQ
  • Baskets & capacities
  • Configuration options
Techn. Data VQ

VQ-82

Art.-No. without exhaust air filtration

26001 (including exhaust air filtration: 26002)

Chamber Volume liters 82
Effective usable volume liters 71
Chamber dimensions Ø x H mm 380 x 756
External dimensions B x T x H mm 670 x 580 x 1010
Loading height mm 870
Required space B x T x H mm 780 x 970 x 1500
Electrical power kW heater 3.6 / pipe heater 0.043
Vacuum pump 0.15
Weight kg 135
Power Supply V AC 400V - 50/60 Hz - 3-phase
Allowed operating temperature °C 140
Selectable operat. temperature °C 105 - 135 (Selection depending on the program)
Warming temperature °C 40 - 60
Max. operating pressure bar 2.2
Timer function no
Sterilization time min. 0 - 419
Locking system Quick release with hand lever
Certification 2014/68/EU, CE 0035
2014/35/EU
2014/30/EU
2011/65/EU
Vessel material SUS304 (stainless steel)
Integrated steam generator No
Vacuum yes
Safety systems Electromagnetic safety lock system, safety valve, low water cut-out, overpressure switch, power cut-off switch, fault indicator (visual and audible)
Documentation SD/HC with a maximum capacity of 32 GB, thermal printer
VQ-Series /
Loading capacities
Basket
Art. No. 56002
Ext. dimensions (Ø x H) in mm 363 x 290
Int. dimensions (Ø x H) in mm 355 x 280
Baskets p. device 2
Configuration options Art. No.
Device including exhaust air filtration 26002
Automatic feed water standard
Integrated printer standard
Flexible temperature sensor standard
Pipe heater for AGAR standard
Air cooling standard
Vacuum standard
SD-Card Slot standard
Practical castors standard

We are happy to assist you!