Technical data | Unit | Safeguard Cyto 900 | Safeguard Cyto 1200 | Safeguard Cyto 1500 | Safeguard Cyto 1800 |
Articlenumber | on request | on request | on request | on request | |
External dimensions excl support, WxDxH | mm | 1003x799x1843 | 1303 x 799 x 1843 | 1603 x 799 x 1843 | 1907 x 799 x 1843 |
External dimensions incl support, table top at 800 mm, WxDxH |
mm | 1003 x 799 x 2053 | 1303 x 799 x 2053 | 1603 x 799 x 2053 | 1907 x 799 x 2053 |
Internal dimensions, WxDxH | mm | 900 x 520(625) x 700 | 1200 x 520(625) x 700 | 1500 x 520(625) x 700 | 1800 x 520(625) x 700 |
Front window opening | mm | 200 ( adjustable 160 - 330) | |||
Work height, basic support stand | mm | 750 - 1050 | |||
Minimum door way | mm | 800 | |||
Weight | |||||
Equipment | kg | 210 | 240 | 270 | 300 |
Air velocity | |||||
Vertical flow | m/s | 0,28 ( adjustable 0,25 - 0,53) | |||
Deviation | ±% | <8 | |||
Noise | |||||
Noise level, measured according to EN12469 | dB(A) | 53 | 54 | 55 | 56 |
Ventilation | |||||
Down flow rate | m3/h | 510 | 680 | 840 | 1010 |
Exhaust flow rate | m3/h | 290 | 390 | 480 | 580 |
Filter technology | |||||
Main/Exhaust | Typ | HEPA H14 EN1822, 99,999% bei 0,3 μm particle size | |||
Light | |||||
LED | Lux | 0-2000, dimmable | |||
Electrical data | |||||
Voltage frequency | V/Hz | 220-240/ 50-60 | |||
Power consumption | w | 90 | 100 | 130 | 140 |