| Technical data | Unit | Safeguard Cyto 900 | Safeguard Cyto 1200 | Safeguard Cyto 1500 | Safeguard Cyto 1800 |
| Articlenumber | on request | on request | on request | on request | |
| External dimensions excl support, WxDxH | mm | 1003x799x1843 | 1303 x 799 x 1843 | 1603 x 799 x 1843 | 1907 x 799 x 1843 |
| External dimensions incl support, table top at 800 mm, WxDxH |
mm | 1003 x 799 x 2053 | 1303 x 799 x 2053 | 1603 x 799 x 2053 | 1907 x 799 x 2053 |
| Internal dimensions, WxDxH | mm | 900 x 520(625) x 700 | 1200 x 520(625) x 700 | 1500 x 520(625) x 700 | 1800 x 520(625) x 700 |
| Front window opening | mm | 200 ( adjustable 160 - 330) | |||
| Work height, basic support stand | mm | 750 - 1050 | |||
| Minimum door way | mm | 800 | |||
| Weight | |||||
| Equipment | kg | 210 | 240 | 270 | 300 |
| Air velocity | |||||
| Vertical flow | m/s | 0,28 ( adjustable 0,25 - 0,53) | |||
| Deviation | ±% | <8 | |||
| Noise | |||||
| Noise level, measured according to EN12469 | dB(A) | 53 | 54 | 55 | 56 |
| Ventilation | |||||
| Down flow rate | m3/h | 510 | 680 | 840 | 1010 |
| Exhaust flow rate | m3/h | 290 | 390 | 480 | 580 |
| Filter technology | |||||
| Main/Exhaust | Typ | HEPA H14 EN1822, 99,999% bei 0,3 μm particle size | |||
| Light | |||||
| LED | Lux | 0-2000, dimmable | |||
| Electrical data | |||||
| Voltage frequency | V/Hz | 220-240/ 50-60 | |||
| Power consumption | w | 90 | 100 | 130 | 140 |